Up Next
Go up to 6 Getting good extra-focal images.
Go forward to 6.2 Special considerations for astronomical telescopes.

6.1 General considerations

The program ef , is designed to compute optical wavefronts from extra-focal images. The algorithms used in the program are based upon the fact that an aberration in the optical wavefront will change the shape and intensity distribution within a de-focused pupil image. More explicitly we use the optical transport equation:
Where I(x,y,z) is the irradiance distribution along the beam and W(x,y,z) is the Wavefront surface referenced to the z (propagation) direction.

The quality of the results obtained is dependent on both the characteristics of your optical system, and the care with which you take measurements.

 

The basic measurements needed by the program are the intensity distribution in two extra-focal planes. In the above figure We show diagrammatically the placement of extra-focal planes around a focal plane. Note that the distance to each extra-focal plane (from the image) plane is the same, this is not essential, but the closer you come to this ideal situation, the better your wavefront measurements are likely to be.

There are many possible ways to obtain these extra-focal images, the most straightforward being simply to translate your camera along the optical axis of your imaging system. With some optical systems, you may be able to adjust the focus of the optics. However re-focusing may induce different aberrations from your optical system at each extra-focal plane and thus adversely affect your wavefront measurement. A common case of this, is re-focusing a cassegrain telescope, which introduces variable spherical aberration, and possibly coma if the focus mechanism is not accurately aligned.

In the following sections we will discuss the issues associated with taking extra-focal images. We have split the discussion into several sections, but none of the considerations is completely orthogonal to the others. It is often beneficial to reduce data taken under several different conditions to quantify the wavefront measurement error.

6.1.1 Definition of the optical pupil.

Extra-focal images should be placed symmetrically either side of the stop in your optical system. The program makes that assumption that the pupil stop is close to infinite conjugate to the image plane. In the current version of the program it is assumed that the optical stop is circularly symmetric. If you wish to obtain wavefronts from systems lacking circular symmetry please discuss your needs with Laplacian Optics Inc. .

Some optical systems have multiple planes which nearly define the pupil stop. This can lead to non circular symmetry if the optical system is mis-aligned. A good example is an infrared astronomical telescope, where the secondary mirror normally defines the optical stop. However the primary mirror edge is only slightly beyond the secondary edge. Under these circumstances a telescope mis-alignment can result in the optical stop being defined by the secondary mirror in some places and the primary mirror in others. The mis-alignment will cause the effective optical stop to be non circularly symmetric. The retrieved wavefront will therefore be in error, the main effect being to reduce the accuracy of low order terms such as astigmatism and coma. The best course of action in this case would be to use the measured wavefronts to calculate the adjustments needed to align the optical system. If the mis-alignment is large, several iterations might be required in order to achieve adequate alignment,

6.1.2 The size of the extra-focal image.

Depending upon your optical system, the physical size of the extra-focal image may be related to the extra-focal distance. We generally recommend using a lens system to take extra-focal images, partly because such a system allows one to decouple the pupil image size from the extra-focal distance, please contact Laplacian Optics Inc. . for more information.

The size of the reconstructed wavefront is not directly related extra-focal image diameter, but is set independently as a part of the data reduction process. In general the resolution on the reconstructed wavefront is substantially lower than that of the extra-focal images.

There are two basic factors which must be considered when attempting to set the size of the extra-focal images:

6.1.2.1 Program considerations.

Due to the transformation of the extra-focal images that takes place whilst this program is running, it is wise to have several times more pixels across the extra-focal image, than is required in the wavefront. One should on no account try to reconstruct a wavefront of higher resolution than the extra-focal image size. We recommend about 100 to 200 pixels across the diameter of an extra-focal image. The program will work with smaller images, but will tend to give less accurate results. Images larger than the recommended size will work well, but may slow down the data-reduction process.

6.1.2.2 Sampling.

As with all imaging applications one would like to ensure that the extra-focal images are properly sampled. However depending upon the optical configuration this may not be feasible. Obviously any under sampling of the extra-focal images will potentially introduce aliasing errors which will reduce the accuracy of the wavefront reconstruction. Most optical surfaces contain little high frequency aberration so the loss of accuracy due to aliasing is not usually significant. Hence with most optical systems and accuracy requirements, under sampling is likely to produce acceptable results. As a general rule it is recommended that you record images at several different extra-focal distances and/or image samplings. Any large differences in the wavefront derived from images with different sampling or extra-focal distances indicates a problem.

6.1.3 The extra-focal distance.

The program ef core algorithm is based on the fact that wavefront deformations produce brightness variations within the extra-focal pupil image which are proportional to the Laplacian of the wavefront. This relationship is only approximately linear, and most of the work performed by ef is to apply corrections for non-linearity. The more severe the non-linearity the more error is generated in the reconstructed wavefront.

The main trade-offs to be considered when choosing an extra-focal distance is the following:

In practice the trade-off between SNR and linearity is usually the most important. The size of the signal scales as:
sig = w*f*f/l
Where w is the magnitude of the wavefront error, f is the f-ratio of the optical system you are measuring, and l is the extra-focal distance you are using. A good heuristic to employ is that if the curvature signal (intensity difference inside extra-focal image) significantly exceeds 25%, you need to increase the extra-focal distance to avoid excessive non-linearity. The program will still work with larger signal strengths, but will give reduced accuracy. Note the signal may be observed by loading two extra-focal image and hitting the reset followed by the iterate buttons in the Reduce :: Run reduction manually menu sub-window. After the iteration has finished the signal will be displayed in the left image window.

6.1.4 The image SNR.

We recommend that data for this program always be taken with a slow scan CCD with at least 10 bits of dynamic range. With even an inexpensive (less than US$1000) amateur grade slow scan CCD the image SNR will be dominated by photon noise when the pixel intensity reaches more than 10% of full well. We recommend exposing extra-focal images to give a peak intensity of between 10% and 50% of full well, and using dark/background and flat field images if available. In our experience most commercial grade frame grabber/video camera combinations give only 5 to 6 bits of dynamic range which is barely adequate for taking reasonable extra-focal images. Worse still, most TV cameras have a nonlinear transfer function or gamma correction, which seriously compromises the accuracy of any extra-focal images that you record. Lastly the automatic gain control found on some video cameras can change the image gain within a single image thus distorting the image brightness distribution, particularly when imaging a very contrasty object, as the extra-focal image should be. If you do use a frame grabber to acquire images, you should make the pupil images as large as possible to average the signal over many pixels, turn off automatic gain control, and average several image frames in software (or using a hardware 16 bit accumulator) to improve SNR.

6.1.5 Extraneous aberrations.

Under most situations it should be possible to build wavefront sensor optics which do not introduce significant aberrations into the measures wavefront. This is possible, because the image quality required in the extra-focal image is quite modest. The most difficult problem with most measuring systems is to control the effects of seeing within the system being measured. The curvature sensing approach as implemented by the ef program allows two approaches to controlling this source of error.

6.1.5.1 Short exposure measurements.

As with traditional measurement systems, one can take short exposure images to freeze any turbulence aberrations, then average several measurements to reduce the effect of these time varying aberrations. To do this with the curvature system requires setting up the wavefront sensor so that it can record both extra-focal images simultaneously.

6.1.5.2 Long exposure measurements.

The preferred approach to eliminating turbulent seeing effects is to integrate each extra-focal image long enough to average out the effects of seeing. The amount of time required to to this depends on the seeing, but is typically of the order of 1 minute or longer. For the special case of atmospheric errors see later.

A neutral density or narrow band filter can be used directly in front of the camera to reduce problems with background light. Small variations in filter should have little impact on the wavefront measurement, provided both extra-focal images fall on the same set of camera pixels (pass through the same piece of filter), and a lens system is used so that images need not be inverted.

When seeing is a consideration we recommend taking several sets of data so that you can compute directly the random errors on the measured wavefront.

6.1.6 Things that can go wrong.

One of the main advantages of the curvature wavefront technique, is its simplicity and low cost of implementation. A related drawback to the method is that it can be adversely affected by various pathological optical problems. Such problems can usually be mitigated by either adding additional optics to the test system, or by suitable configuration of the software. The main purpose of this section is to make you aware of potential problems to enable you to best apply this software.

6.1.6.1 Turned edge on wavefront.

One of the most common optical aberrations encountered in optical fabrication is errors on the edges of an optical component. In operation this can lead to light scatter from the edge of the optical pupil. Since the program uses the shape of the pupil to derive boundary conditions for solving the Poisson equation, light scatter at the pupil circumference can prove problematical. The most noticeable effect of this problem is difficulty in accurately finding wavefront Zernike terms with zero curvature. Most commonly this produces inaccurate measurements for the astigmatism terms.

The first thing you should do under these circumstances is to increase the extra-focal distance. You should try to get far enough from focus to guarantee that the light from the rolled edge is not in the caustic Zone. This will reduce the SNR for all other aberrations, but there is no other choice. If you cannot increase the extra-focal distance sufficiently you may try the reduction sequence rolled available under the Edit :: Select reduction sequence is tuned to give better results with a moderately rolled wavefront edge. Be warned however that this does not always work, and even if it does, the reconstructed wavefronts will be less reliable.

For severe cases of turned edges, we would recommend excluding the damaged part of the wavefront by using a physical pupil stop. This is likely to improve the imaging performance of your optics as well!

6.1.6.2 Chromatic aberration.

In systems where significant chromatic aberrations may be present, you should not use wide band imaging for the extra-focal images. Typically chromatic aberration will produce a signal which looks very similar to spherical aberration. Ideal in this case would be to measure the extra-focal images using several narrow band filters, which would allow determination of the systems chromatic behavior.

6.1.6.3 Speckles and fringing.

When using a monochromatic source to measure extra-focal images, you may observe speckle or low-level fringing in the extra-focal image plane, due to reflections or scattering within the optical system. If the structures are properly sampled, and fixed with respect to the extra-focal image coordinates, they will theoretically, have no effect when measuring small aberrations. In practice these conditions are unlikely be met, so you should treat speckle and fringing as a source of measurement noise. It is probably unwise to under-sample fringing structures, since this could give large low frequency errors, under-sampling speckle will be less severe.

A possible optical solution is to introduce time varying random aberrations (eg seeing) and allow these to average out some of the speckle (and possibly fringing).


Laplacian Optics Inc.     Email:   laplace@laplacian.com

Up Next